Synthesisers (RF Immunity)
The Laplace Synthesisers cover both conducted and radiated RF immunity test requirements. The frequency coverage extends from 100KHz to 230MHz for conducted tests (RFIC-4-6) and 30MHz – 6GHz for radiated tests (RF3000/RF6000). They are all controlled from the users PC via a USB link, and powerful Windows software is included to enable the user to control all the parameters of the test, to monitor the test in progress, and to view the results.

Key Points
- Creates the signal to be applied to the product
- Controls the intensity of that signal
- Monitors the condition of the EUT
- Acts as the interface between a PC and the test
- Includes PC (Windows) software to enable very simple operation
- Complies with IEC61000-4-3 (RF3000/RF6000)
- Complies with IEC61000-4-6 (RFIC-4-6)
Software Details
The RFSynth software will run under any version of Windows from XP to Win10. It allows the following parameters to be set:
- Start frequency
- Stop frequency
- Frequency step (%)
- Stress level (V/m)
- Modulation (AM and pulsed)
- Dwell time
- EUT monitoring (ignore/stop/continue)
- EUT prompt output
It includes facilities to load calibration data for the test cell, field probe or CDN as appropriate. Outputs include:
- Display of target and actual stress level vs frequency
- EUT monitor
- System status
- Hardcopy via Windows printers
- Results file storage to disk and subsequent retrieval
Product Specification
Parameter | RFIC-4-6 | RF3000 | RF6000 |
Start frequency | 100KHz | 30MHz | 30MHz |
End frequency | 300MHz | 3GHz | 6GHz |
Control feedback | RF (from CDN cal. adaptor) DC (from E series CDN) Current Clamp |
DC from Laplacell RS232 from sensor |
|
Compatible with | CDN EM Clamp BCI |
Laplacell Anechoic chamber |
|
Level control mode | Pre-calibrate scan (conventional mode) Closed loop (with E series CDN) |
Pre-calibrate scan or closed loop |
Common to all synthesisers:
Frequency step | 0.1% to 10% of current value | |
Output Level | -60dBm to 0dBm (carrier signal) | |
Output level Indication
Feedback level indication |
Bargraph display | |
THD (worst case) | 10% | |
Modulation | 1KHz sine, 80% AM modulation.
217Hz, 200Hz, 18Hz, 10Hz & 1Hz pulsed. 100% level |
|
Control Mode | Open loop (Pre-scanned) or closed loop. (Software selectable) | |
EUT status | Volt free contact relay input | |
Fault modes | Stop, pause, continue | |
EUT prompt | 4 pole c/o volt free contacts | |
Modes | Pulsed, Continuous, off |